تجاوز إلى المحتوى الرئيسي

Proper Orthogonal Decomposition Analysis of the Flow Downstream of a Dysfunctional Bileaflet Mechanical Aortic Valve

ملخص البحث

Abstract

Purpose

Aortic valve replacement remains the only viable solution for symptomatic patients with severe aortic valve stenosis. Despite their improved design and long history of successful operation, bileaflet mechanical heart valves are still associated with post-operative complications leading to valve dysfunction. Thus, the flow dynamics can be highly disturbed downstream of the dysfunctional valve.

Methods

In this in vitro study, the flow dynamics downstream of healthy and dysfunctional bileaflet mechanical heart valves have been investigated using particle image velocimetry measurements. Proper orthogonal decomposition of the velocity field has been performed in order to explore the coherent flow features in the ascending aorta in the presence of a dysfunctional bileaflet mechanical heart valve.

Results

The ability of proper orthogonal decomposition derived metrics to differentiate between heathy and dysfunctional cases is reported. Moreover, reduced-order modeling using proper orthogonal decomposition is thoroughly investigated not only for the velocity field but also for higher order flow characteristics such as time average wall shear stress, oscillatory shear index and viscous energy dissipation.

Conclusion

Considering these results, proper orthogonal decomposition can provide a rapid binary classifier to evaluate if the bileaflet mechanical valve deviates from its normal operating conditions. Moreover, the study shows that the size of the reduced-order model depends on which flow parameter is required to be reconstructed.

مؤلف البحث
A Darwish, G Di Labbio, W Saleh, L Kadem‏
تاريخ البحث
مجلة البحث
Cardiovascular Engineering and Technology
مؤلف البحث
صفحات البحث
14
الناشر
Springer
تصنيف البحث
International Journal
عدد البحث
12
موقع البحث
https://link.springer.com/article/10.1007/s13239-021-00519-w
سنة البحث
2021